Chattanooga Fluidotherapy

Chattanooga Fluidotherapy Extremity Unit

The Chattanooga Fluidotherapy Extremity Unit is a dry-heat cabinet that fluidises a bed of Cellex, Chattanooga's processed natural-cellulose medium. A blower drives heated air up through a distributor plate into the chamber until the particle bed behaves like a low-viscosity fluid, and the extremity goes in through a sleeved opening. Time, temperature and air speed are set independently on the front panel, and the unit runs in continuous or pulsed mode.

Chattanooga builds the range in two cabinets, and the single extremity unit, part FLU110D, is 34 x 11.5 x 33 in. and weighs 70 lb less medium. The double extremity unit, part FLU115D, is 34 x 18.5 x 33 in., weighs 60 lb without medium and is configured for two extremities. Both ship at 100 lb. Operating temperature is 88 F to 125 F (31 C to 52 C) depending on the air speed selected, treatment time runs 1 to 99 minutes or continuous, air speed runs 0 to 100 percent in 5 percent increments, and pulse is adjustable from OFF to 6 seconds ON and OFF.

The number that governs running cost is the medium charge. The single unit holds approximately 30 lb (13.6 kg) of Cellex and the double approximately 40 lb (18.1 kg), with 30 lb of Cellex included when either unit ships. Cellex comes in 10 lb containers, so a double takes four to charge and a single takes three, and a full medium change falls due every six months. Both units are Class I, Type B, certified to UL 60601-1, and run on 120 V, 50/60 Hz at 11 A. The operator sets time, temperature and air speed from the front panel with no software layer in between. Ownership cost after the medium is the service schedule: inlet filters cleaned daily, sleeves laundered weekly, and a certified technician in every quarter for cavity cleaning, intake filter replacement and functional testing.

Chattanooga Fluidotherapy single extremity dry heat unit with sleeved treatment opening

Operating temperature

88-125 F

31 to 52 C, by air speed

Air speed

0-100 pct

5 percent increments

Treatment time

1-99 min

Plus continuous; pulse to 6 sec

Cellex charge

30 or 40 lb

Single 30 lb; double 40 lb

Two units: single FLU110D, double FLU115D

Cellex medium MED0001 in 10 lb containers

Class I, Type B; UL 60601-1

Scheduled maintenance every 6 months

The machine

Heated air turns Cellex into a fluid

A blower drives heated air up through a distributor plate until the bed of Cellex behaves like a fluid, and the front panel carries three settings that run independently of each other: treatment time, temperature and air speed. The bed runs steady or pulsed, and the same panel sits on the single cabinet and the double cabinet alike.

01

A fluidised bed of Cellex

Heated air is driven up through a distributor plate into a chamber packed with Cellex, a medium Chattanooga makes from natural cellulose, environmentally safe and easily disposable. Only Cellex dry heat medium goes in these units, and the distributor plate is one of the parts on the six-month replacement schedule.

Natural cellulose

02

Three settings you move one at a time

Three adjustments that move independently of each other on the front panel. Treatment time runs 1 to 99 minutes or continuous, air speed runs 0 to 100 percent in 5 percent increments, and the cabinet holds 88 F to 125 F with 115 F to 118 F at 50 percent air speed as the reference operating point.

115-118 F reference

03

Continuous and pulsed modes

Pulse is adjustable from OFF to 6 seconds ON and OFF, so the bed can be run steady or cycled through the treatment time. Turn the unit OFF before a patient is positioned in it or taken out of it, a habit worth setting on the day the machine arrives.

OFF to 6 sec

04

Single or double cabinet

FLU110D is 11.5 in. wide across a 34 in. length and a 33 in. height; FLU115D is 18.5 in. wide over the same length and height and is configured for two extremities. Cabinet width is the sizing figure to plan a room around, since both cabinets are floor-standing and neither changes footprint in use.

11.5 or 18.5 in.

Ownership

Filters get cleaned every day it runs

Fluidotherapy carries a heavier scheduled-maintenance load than any other heat modality Chattanooga sells, from a daily filter clean to a full change of the Cellex medium twice a year, and the cost of that schedule belongs in the purchase decision.

01

Cellex every six months

A full change of the Cellex medium falls due at the six-month service, which is 30 lb on a single unit and 40 lb on a double, twice a year.

02

Daily filter and sleeve checks

The maintenance schedule calls for cleaning the inlet filters and inspecting sleeve condition daily.

03

Weekly sleeve laundering

All sleeves are laundered weekly in mild antibacterial detergent, and the top and end sleeves are removed and inspected.

04

Quarterly technician service

Internal cavity inspection and cleaning, intake filter replacement, full functional and performance tests, distributor replacement if needed.

05

Annual calibration

An annual item on the required maintenance schedule, alongside blower motor and distributor replacement at six months.

06

120 V, 50/60 Hz, 11 A

Electrical supply for the single and the double unit alike, at Class I, Type B, certified to UL 60601-1.

07

30 lb of Cellex in the carton

Either cabinet ships with 30 lb of Cellex included, which is a full charge for the single unit and one 10 lb container short of a full double.

08

Warranty by part

One year from original purchase, 90 days on accessories, 180 days on motors and distributors.

In detail

Width and Cellex charge separate the two cabinets

Chattanooga Fluidotherapy double extremity dry heat unit with the wider cabinet

Single or double

Seven inches of cabinet and 10 lb of medium apart

The two units share a 34 in. length, a 33 in. height, a 100 lb shipping weight and every control specification on the panel. What separates them is cabinet width, 11.5 in. against 18.5 in., the configuration for one extremity or two, and the medium charge of approximately 30 lb against approximately 40 lb of Cellex. Cabinet weight without medium is 70 lb on the single FLU110D and 60 lb on the double FLU115D, so both units land at roughly 100 lb once they are charged and standing on a clinic floor.

  • FLU110D single: 34 x 11.5 x 33 in., 70 lb less medium, approximately 30 lb of Cellex
  • FLU115D double: 34 x 18.5 x 33 in., 60 lb without medium, approximately 40 lb of Cellex
  • Both: 120 V 50/60 Hz 11 A, shipping weight 100 lb, Class I Type B, UL 60601-1
  • Both: 1 to 99 minutes and continuous, air speed 0 to 100 percent in 5 percent increments
  • Cabinet width is the sizing dimension, 11.5 in. on the single and 18.5 in. on the double
11.5 or 18.5 in. 30 or 40 lb of media 100 lb shipping
Chattanooga Cellex dry heat media container for Fluidotherapy units

Fitment

A full charge takes three or four containers

Cellex dry heat media is Chattanooga part MED0001, sold in 10 lb (5 kg) containers, and it is the only medium that goes in these units. A single FLU110D takes approximately 30 lb, which is three containers; a double FLU115D takes approximately 40 lb, which is four. Either unit ships with 30 lb of Cellex included, so a double needs a fourth container on the same order to reach a full working charge. Budget for six containers a year on a single and eight on a double, because the whole charge is replaced at every six-month service.

  • Cellex dry heat media, part MED0001, 10 lb (5 kg) per container
  • Single FLU110D: approximately 30 lb, which is three containers
  • Double FLU115D: approximately 40 lb, which is four containers
  • A full medium change falls due every six months on the maintenance schedule
  • Only Cellex dry heat medium goes in a Fluidotherapy unit
MED0001 10 lb containers Changed every 6 months

The range

Dry heat against two moist-heat tanks

Three heated-treatment machines from Chattanooga go into a clinic room, and two of them are moist-heat water baths holding HotPacs. Fluidotherapy is the dry-heat cabinet, the only one of the three with a treatment timer and an air speed control on the unit itself.

You're Here

Chattanooga Fluidotherapy single extremity dry heat unit

Fluidotherapy

Cabinet of heated Cellex cellulose media, 88 F to 125 F, air speed 0 to 100%.

  • Single FLU110D, double FLU115D
  • Medium approx 30 lb single; approx 40 lb double
  • Treatment timer 1 to 99 minutes and continuous
  • Air speed 0% to 100% in 5% increments
Currently viewing
Chattanooga Hydrocollator M-4 mobile heating unit

Hydrocollator mobile

Floor-standing tank, 10 to 27 gal by model, supplied with 8 to 24 Standard HotPacs.

  • SS-2, M-2 and M-4 tank sizes
  • Operating temperature 160 to 165 F (71 to 74 C)
  • Filled weight 140 lb SS-2; 190 lb M-2; 340 lb M-4
  • Consumable supplied: 8 Standard HotPacs SS-2; 12 M-2; 24 M-4
Shop the mobile units
Chattanooga Hydrocollator E-2 tabletop heating unit

Hydrocollator tabletop

Benchtop tank, 3 or 8 gal. The E-1 reaches 160 F in 2 hours.

  • E-1 and E-2 bench tanks
  • Operating temperature 160 to 165 F (71 to 74 C)
  • Filled weight 55 lb E-1; 125 lb E-2
  • Consumable supplied: 4 Standard HotPacs E-1; 2 oversize, 3 standard, 1 neck contour E-2
Shop the tabletop units

Questions

What the maintenance schedule actually requires

120 V, 50/60 Hz, 11 A on both the single FLU110D and the double FLU115D, with the units classified Class I, Type B and certified to UL 60601-1. Eleven amps is a real draw for a treatment room, so put the cabinet on an outlet that is not already carrying the rest of the room. Export versions, FLU110DE and FLU115DE, run 210 to 250 VAC at 8 A.

From Chattanooga's manual. The device should not be used for symptomatic relief of pain unless etiology is established or unless a pain syndrome has been diagnosed. It should not be used when cancerous lesions or open wounds are present in the treatment area. The same warning covers patients suspected of carrying serious infectious disease, and disease where it is advisable for general medical purposes to suppress heat or fevers. Precautions apply to patients with heart problems, epilepsy and diabetes, and a medical professional familiar with those precautions should be consulted about any patient who may react to dust and pollen.

Either cabinet ships with 30 lb of Cellex included, and the sleeved treatment opening is built into the cabinet, not bought separately. A single FLU110D holds approximately 30 lb at a full charge, so the included medium fills it on day one. A double FLU115D holds approximately 40 lb, so add one 10 lb container of MED0001 to the order and the second cabinet is ready when it lands. Budget three containers a charge on the single and four on the double from the first six-month change onward.

Approximately 30 lb (13.6 kg) for the single FLU110D and approximately 40 lb (18.1 kg) for the double FLU115D. Either unit ships with 30 lb of Cellex included, so a double is one 10 lb container short of a full charge on delivery. Cellex is part MED0001 in 10 lb containers: three for a single, four for a double.

Every six months, on Chattanooga's required maintenance schedule. That works out at six 10 lb containers a year for a single unit and eight for a double. Blower motor and distributor replacement sit at the same six-month service, so the medium change and the mechanical service are one visit.

Chattanooga sets it out by interval. Daily: clean the inlet filters, inspect sleeve condition. Weekly: launder all sleeves in mild antibacterial detergent, remove and inspect top and end sleeves. Quarterly: internal cavity inspection and cleaning, intake filter replacement, full functional and performance tests, distributor replacement if needed. Every six months: change the Cellex medium, replace the blower motor, replace the distributor. Annually: calibration. Quarterly work is specified for a certified technician.

Chattanooga puts this in a warning: "Failure to perform the Required Maintenance could result in the Cellex medium entering the heat chamber of the unit(s) and cause severe injury to patients as well as smoke damage to the facility and the Fluidotherapy unit(s)." Do not use the unit in the presence of flammable anesthetics, where there is an explosion hazard. Put the quarterly technician visit in the calendar before the machine is delivered.

11.5 in. for the single FLU110D and 18.5 in. for the double FLU115D, both over a 34 in. length and a 33 in. height. Cabinet width is the figure to plan the room from, since the double fits two extremities into the same length and height as the single. If you are sizing the machine to a particular patient population, ask us and we will put the question to Chattanooga against the model on your quote.

88 F to 125 F (31 C to 52 C), dependent on the air speed selected, with 115 F to 118 F at 50 percent air speed as the reference operating point. Air speed is set from 0 to 100 percent in 5 percent increments and treatment time from 1 to 99 minutes or continuous, and pulse runs from OFF to 6 seconds ON and OFF.

The single FLU110D is 70 lb less medium and the double FLU115D is 60 lb without medium, and shipping weight is 100 lb for both. Add the Cellex charge on top, approximately 30 lb in the single and approximately 40 lb in the double, and both cabinets come out at roughly 100 lb in service. Both stand 33 in. tall over a 34 in. length, so these are floor-standing machines and no bench is involved.

No. Use only Cellex dry heat medium in the Fluidotherapy units. Cellex is made from natural cellulose, environmentally safe and easily disposable, and it is part MED0001 in 10 lb containers.

One year from the original purchase date, 90 days on accessories, and 180 days on motors and distributors, with repairs without charge within thirty days on return. It excludes unauthorised repair, misuse, failure to perform the required maintenance, and damage caused by non-certified technicians, which is one more reason to keep the quarterly technician service on the books.

A third model exists, the FLU210D, at 44 x 12.5 x 35.75 in., 140 lb, with a 60 lb medium capacity. The two current US units are the single FLU110D and the double FLU115D, so ask us if the larger cabinet is what your room needs and we will check availability with Chattanooga before you write the order.

Different machines for different work. Fluidotherapy is an extremity cabinet with a timer, a temperature setting and an air speed control, and the limb goes inside it. A Hydrocollator is a 160 to 165 F water bath that keeps HotPacs hot for application anywhere on the body, with no timer and no per-treatment settings on the unit itself. A clinic that wants a set timer for extremity work and hot packs for everywhere else runs one of each, and the two machines have nothing in common on the maintenance schedule.

  • Part numbers: FLU110D single, FLU115D double; export FLU110DE, FLU115DE
  • Operating temperature: 88-125 F (31-52 C), by air speed
  • Reference operating point: 115-118 F at 50 pct air speed
  • Treatment time: 1-99 minutes and continuous
  • Air speed: 0-100 pct in 5 pct increments
  • Pulse mode: OFF to 6 seconds; ON/OFF
  • Dimensions: 34 x 11.5 x 33 in. single; 34 x 18.5 x 33 in. double
  • Product weight: 70 lb (31.7 kg) single less medium; 60 lb (27.2 kg) double without medium
  • Shipping weight: 100 lb (45.4 kg), both units
  • Mains power: 120 V, 50/60 Hz, 11 A
  • Export power: 210-250 VAC, 50/60 Hz, 8 A on the E models
  • Safety class: Class I, Type B
  • Safety tests: UL 60601-1
  • Distributor plate and blower motor: Both replaced at the six-month service
  • Warranty: 1 year from original purchase; 90 days accessories; 180 days motors and distributors, per the manual
  • Units: Single extremity FLU110D; double extremity FLU115D
  • Medium: Cellex dry heat media, part MED0001, 10 lb (5 kg) containers
  • Medium charge: Approx 30 lb single (3 containers); approx 40 lb double (4 containers)
  • Medium change interval: Every 6 months, per the manual's required maintenance schedule
  • Variable adjustments: Time, temperature and air speed, set independently
  • Modes: Continuous and pulsed
  • Included medium: 30 lb of Cellex ships with either cabinet, a full charge on the single
  • Sleeves: Sleeved treatment opening; laundered weekly, top and end sleeves removed and inspected
  • Contraindications: Cancerous lesions or open wounds in the treatment area; suspected serious infectious disease; disease where suppressing heat or fevers is advisable
  • Precautions: Heart problems, epilepsy, diabetes; allergic reaction to dust and pollen
  • Daily and weekly maintenance: Clean inlet filters and inspect sleeves daily; launder sleeves weekly
  • Quarterly maintenance: Certified technician: cavity cleaning, intake filter replacement, performance tests
  • Medium restriction: Only Cellex dry heat medium goes in these units, part MED0001
  • Cabinet width: 11.5 in. single; 18.5 in. double